Test current
From 5 A up to 100 A (True DC)
The test current can be adjusted in steps of 0.2 A from 5 A to 20 A and steps of 1 A from 20 A to 100 A.
Resistance ranges
0.1 μΩ up to 2 mΩ (0.1 μΩ resolution)
2 mΩ up to 300 mΩ (10 μΩ resolution).
Measurement principle
Four-terminal, U/I.
Protections
Protections against overheating, overcurrent and short-circuit.
Programmable test time
Allows to setup the test duration from 15 seconds up to 120 seconds for current test from 5 A up to 100 A
Basic accuracy
± 1 % of reading from 50 μΩ to 300 mΩ.
Auxiliary clamp measurement range
0.1 ADC to 60 ADC.
Auxiliary clamp reading accuracy
1 % + 3 digits.
Advanced features
Digital direct reading of very low resistances in the alphanumerical display, with up to 4½ digits. Very fast and accurate measurements.
Communication protocol
Modbus
Built-in memory
Memory for storing up to 4000 readings organized by records.
Serial data output
USB, for configuration, control and download the stored values
Bluetooth
For configuration and control
Environmental protection
IP54 (with closed lid)
Safety
In accordance with IEC 61010-1
Power supply
Mains: 100-240 V~
Operating temperature
32 °F to 122 °F (0 °C to 50 °C)
Storage temperature
14 °F to 158 °F (-10°C to 70°C)
Humidity range
95% RH (non condensing)
Equipment weight
approx. 24.25 lb (11 kg)
Dimensions
502 x 394 x 190 mm